60W open frame AC-DC medical-grade power supply ACMS61B

This Longxc 60W open frame AC-DC power supply is designed specifically for medical applications, providing high performance, safety, and reliability. Engineered to comply with stringent medical safety standards (IEC 60601-1 3rd edition), it ensures low leakage current and reinforced isolation, making it ideal for use in patient-connected devices and diagnostic equipment.

Request a Quote

PRODUCTS DETAILS

PRODUCT PICTURE

DESCRIPTION

The purpose of the document is to specify the functional requirements of a ACMS61B-120500 medical switching power

supply.

/MODEL NAME RULE

ACMS61B-aaabbb

-ACMS=Single-channel medical switching power supply

-61=Name of the series

-aaa= "120"stands for the output voltage."120" means 12.0V.

-bbb="500"stands for the output voltage."500" means 5.0A.

AC Leakage Current

Normal touch current less than 0.1mA at 264Vac input.

Single fault touch current less than 0.5mA at 264Vac input.

Normal earth leakage current less than 5mA at 264Vac input.

Single fault earth leakage current less than 10mA at 264Vac input.

PROTECTION FUNCTION

6.1 Short Circuit Test

The power supply will be auto recovering when the short circuit faults removed.

6.2 Over Current Protection

When the output current reaches 110%~200% of the rated current, the power supply will be protected. After

the over current fault is removed, the power supply will be automatically recovering.

SAFETY REQUIREMENT

Safety: accord with.

The power supply is designed for indoor use to meet IEC 60601-1/UL60601-1and EN 60601-1standards and for

the following safety regulations.

Dielectric Strength Hi-Pot

The primary to secondary: 4000Vac 5mA for 60 seconds

The primary to earth: 1500Vac 5mA for 60 seconds

The secondary to earth:500Vdc 5mA for 60 seconds

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours

Leave YourMessage

Thank you! We will reply you via email within 24 hours.

Return